Research Article | | Volume 14 Special Issue 1 (Jan-April, 2025) | Pages 392 - 396

Influence of Various Implant-Abutment Connection Designs on Microleakage and Bacterial Penetration

 ,
 ,
 ,
 ,
 ,
1
Department of Prosthodontics, Hitech Dental College, Research scholar, Institute of dental science S”o”A university Bhubaneswar, Odisha- 751025, India
2
Department of Prosthodontics, Faculty of Dentistry, Taif University, Taif, 21944, Saudi Arabia
3
Department of Oral and Maxillofacial Surgery, RKDF Dental College and Research Center, Bhopal, 462026, India
4
Restorative Dental Sciences, Faculty of Dentistry, Taif University, Taif, 21944, Saudi Arabia
5
Department of Oral and Maxillofacial Surgery, Inderprastha Dental College and Hospital, Sahibabad, Ghaziabad , Uttar Pradesh, 201010, India
6
Department of oral Medicine and Radiology, ⁠Kalinga Institute of Dental Sciences , Bhubaneswar, India
Under a Creative Commons license
Open Access
Received
Feb. 22, 2025
Revised
March 18, 2025
Accepted
April 10, 2025
Published
April 27, 2025

Abstract

Objectives: The connection between implant and abutment establishes essential conditions for lasting success of dental implant therapy. Bacterial colonization at the implant-abutment connection causes microleakage which subsequently leads to peri-implant diseases as well as eventual implant failure. The Morse taper together with internal hexagon and external hexagon represent three implant-abutment connection designs which seek to decrease bacterial penetration. The present study examines how variations in implant-abutment connections affect both microleakage and bacterial infiltration between components. Methods: This in vitro research examined 60 titanium dental implants through three distinct connection design categories consisting of Morse taper with 20 units and internal hexagon with 20 units and external hexagon with 20 units. A bacterial exposure with Streptococcus mutans occurred on each implant-abutment unit before placing them in a 37°C environment during a seven-day period. Laboratory examination of bacterial penetration included RT-PCR tests and evaluation of bacterial cultures. A spectrophotometric analysis of dye penetration technique was employed to measure microleakage between groups. The statistical analysis employed one-way ANOVA for data assessment while post-hoc Tukey’s test provided supplemental evaluations at a p value below 0.05. Results: The Morse taper connection group demonstrated better microleakage resistance through its 0.25±0.05 mm readings when compared to the 0.67±0.08 mm readings from internal hexagon and 1.12±0.10 mm readings from external hexagon groups (p<0.05). Bacterial penetration occurred in 10% of Morse taper implants whereas 45% of internal hexagon implants presented bacterial penetration and 75% of external hexagon implants showed bacterial penetration. The Morse taper connection provided advanced sealing ability that reduced both microleakage and bacterial passage through the connection. Conclusion: Design specifications for implant-abutment connections determine how well bacterial agents and fluid can pass through the connection. Morse taper connections show enhanced sealing capabilities which lowers the chance of bacterial infections during the procedure. Clinical application of Morse taper implants should be favored because research shows they help extend implant life and decrease peri-implant bacterial infections. The research demonstrates that Morse taper connections decrease peri-implant diseases effectively because they block bacterial penetration. The clinical use of Morse taper implants should be considered when sustaining implant longevity together with controlling infection rate is essential.

Keywords
Implant-abutment connection, Microleakage, Bacterial penetration, Morse taper, Internal hexagon, External hexagon, Peri-implant disease

INTRODUCTION

The dental implant procedure functions as a widely recognized teeth replacement method which provides better oral functionality and better appearance options. Implant longevity depends heavily on implant-abutment connections since these elements maintain the structure and block bacterial entry into the system [1]. Phageflow of bacteria at the implant-abutment interface leads to peri-implant inflammation as well as bone loss that culminates in implant failure [2,3].

 

Research teams have created numerous implant-abutment connection types to achieve better sealing functions and reduced microleakage. Research shows that the Morse taper connection delivers excellent bacterial sealing through its cone shape that locks together better and creates minimal microgap measurements [4,5]. Internal and external hexagonal connections show higher microleakage potential because both their structural design and their tendency to move from abutment manipulation [6,7]. Microgaps found in these systems serve as bacterial reservoirs that lead to the development of peri-implant diseases according to research studies [8,9].

 

The implant-abutment interface bacterial infiltration can be evaluated through three in vitro techniques which include bacterial culture and real-time polymerase chain reaction (RT-PCR) and spectrophotometric dye penetration analysis [10,11]. The tested methods allow researchers to measure how well various connection approaches prevent bacterial infiltration into implant systems. The identification of design-to-bacterial microleakage relationships in implant-abutment connections provides essential data for developing implant success-rate enhancing and peri-implant complication minimizing methods.

 

The research examines how different implant-abutment interfaces including Morse taper along with internal hexagon and external hexagon affect bacterial penetration and microleakage formation. The examination of these systems seeks to identify the most optimal bacterial sealing properties which will enhance the long-term achievement of dental implants. Medically speaking bacterial microleakage at the implant-abutment junction requires reduction because it extends implant durability and fights peri-implantitis particularly in high-risk patient groups which include diabetic and smoking individuals. The study boundaries excluded analysis of recent surface treatment innovations including nanostructured coatings and antibacterial modifications that demonstrate promise in improved implant integration and bacterial resistance. This research examined the microleakage levels and bacterial penetration within three implant-abutment connection types namely Morse taper, internal hexagon and external hexagon. The experiment analyzed microleakage through spectrophotometric dye assessment as its main outcome and assessed bacterial penetration by combining RT-PCR and CFU quantification techniques. The Morse taper connection should show better sealing abilities than hexagonal devices thus promoting longer implant survival paired with lesser peri-implant tissue disorders.

METHODS

A laboratory investigation examined microleakage alongside bacterial penetration in implant-abutment connection designs under test conditions. A research involving three implant-abutment connection types used 60 titanium dental implants that were divided into Morse taper (n = 20), internal hexagon (n = 20), and external hexagon (n = 20). All implants originated from one producer to guarantee a consistent product material composition together with surface characteristics. Sample size estimation was performed using G*Power software with an effect size of 0.8, power of 0.9, and α = 0.05, which determined a minimum of 18 samples per group. We included 20 samples per group to account for potential losses. Group assignment was randomized using a computer-generated sequence to minimize allocation bias. Surface roughness of all implants was standardized and verified using profilometry prior to grouping to ensure comparability across all samples.

 

Each implant-abutment combination received ultrasonic distilled water treatment lasting five minutes before undergoing 15-minute sterility-enhancing autoclaving at 121°C. The abutment installation onto implants occurred according to manufacture guidelines through the application of a calibrated torque wrench. The specimens spent 24 hours at 37°C under artificial saliva solution for intraoral environment simulation. Sterility was ensured by maintaining aseptic conditions throughout the experiment. All handling of implants and bacterial inoculation procedures were performed in a Class II biosafety cabinet under sterile conditions using autoclaved instruments and consumables.

 

A 0.5 McFarland concentration of Streptococcus mutans bacteria in brain-heart infusion broth served as the reference standard to prepare 10 µL bacterial suspension which received injection into implant-abutment interfaces through a micropipette. Bacterial penetration and biofilm formation was allowed by the samples through an incubation cycle at 37°C within an anaerobic chamber for seven days. To reduce methodological error, all RT-PCR procedures were carried out using validated commercial kits with internal controls. Spectrophotometric readings were conducted in triplicate to ensure reproducibility, and operator calibration was done to minimize intra-observer variability.

 

The evaluation of microleakage occurred through spectrophotometric dye penetration measurements. The researchers placed each implant-abutment combination inside 1% methylene blue dye solution for 48 hours. The implants were cut in half after rinsing in distilled water so the microscope could measure the amount of dye that penetrated through stereomicroscopic observation. The measurement of microleakage occurred in units of millimeters.

 

Calculations using RT-PCR and bacterial cultures determined whether bacteria had penetrated the implant-abutment connection. The inner parts of implants underwent DNA extraction through a commercial bacterial DNA extraction kit before S. mutans-specific primer-based amplification procedures for RT-PCR analysis. BHI agar plates with 24 hours (37°C) incubation following paper point placement into sterile implant-abutment interface allowed counting of colony-forming units (CFUs).

 

The researchers conducted their analysis through SPSS (version 26). The analysis included One-way ANOVA to evaluate microleakage data between the three groups. Post-hoc Tukey’s test determined specific inter-group pair comparisons. A chi-square analysis determined whether bacteria penetrated through the list examined areas. All reported data reached statistical significance with a p value of less than 0.05.

RESULTS

Microleakage Analysis

Table 1 contains the recorded mean microleakage values from the three implant-abutment connection designs. The Morse taper connection achieved the minimum microleakage value of 0.25±0.05 mm compared to 0.67±0.08 mm for the internal hexagon and 1.12±0.10 mm for the external hexagon group. Research findings demonstrated that the three experimental groups presented different microleakage values since statistical significance exceeded 0.05 (p<0.05). Patient treatment benefits from the Morse taper connection as it produces better sealing and smaller microleakage than other implant design methods (Table 1).

 

Table 1: Mean Microleakage (mm) in Different Implant-Abutment Connection Designs

Implant-Abutment Connection

Mean Microleakage (mm)±SD

Morse Taper

0.25±0.05

Internal Hexagon

0.67±0.08

External Hexagon

1.12±0.10

p-value

< 0.05

 

 

The significant difference in microleakage values suggests that Morse taper provides a more hermetic seal at the implant-abutment interface, which may directly translate into improved clinical outcomes by minimizing early bacterial colonization.

 

Bacterial Penetration

The bacterial penetration rates for each implant-abutment connection type are summarized in Table 2. Bacterial infiltration was detected in 10% of the Morse taper group, 45% of the internal hexagon group, and 75% of the external hexagon group (p<0.05). These results suggest that the Morse taper connection is more effective in preventing bacterial contamination compared to the hexagonal connection types (Table 2).

 

 

Table 2: Bacterial Penetration Rate in Different Implant-Abutment Connection Designs

Implant-Abutment Connection

Bacterial Penetration (%)

Morse Taper

10%

Internal Hexagon

45%

External Hexagon

75%

p-value

< 0.05

 

 

These findings have important clinical implications, particularly in reducing the incidence of peri-implant infections in susceptible patient populations by selecting implant systems with lower microbial penetration.

 

Bacterial Load Quantification

The bacterial load inside the implant-abutment interface was assessed using CFU counting and RT-PCR analysis (Table 3). The Morse taper connection showed the lowest bacterial load (2.1×10³ CFU/mL), while the internal and external hexagon connections exhibited significantly higher bacterial contamination (5.8×10³ CFU/mL and 9.4×10³ CFU/mL, respectively). These differences were statistically significant (p<0.05).

 

Table 3: Bacterial Load (CFU/mL) in Different Implant-Abutment Connection Designs

Implant-Abutment Connection

Bacterial Load (CFU/mL)±SD

Morse Taper

2.1×10³±0.5×10³

Internal Hexagon

5.8×10³±1.2×10³

External Hexagon

9.4×10³±1.8×10³

p-value

< 0.05

 

 

The lower bacterial load observed in the Morse taper group supports its effectiveness in resisting biofilm formation, a key contributor to implant complications such as peri-implant mucositis and peri-implantitis.

 

Statistical Findings

One-way ANOVA analysis revealed a significant difference in microleakage, bacterial penetration, and bacterial load across all groups (p<0.05). Post-hoc Tukey’s test confirmed that the Morse taper group had significantly lower microleakage and bacterial penetration compared to the other groups (p<0.05).

 

These results highlight the superior bacterial sealing ability of the Morse taper connection, which could contribute to reducing the risk of peri-implant complications.

DISCUSSION

Designs of implant-abutment connections serve dual functions by establishing implant stability and blocking bacterial penetration that leads to peri-implant diseases. This research study confirmed that Morse taper connections perform better than internal and external hexagonal connections for sealing purposes when analyzing implant-abutment microleakage and bacterial penetration. Studies in the field proved that conical implant-abutment connections decrease bacterial penetration because of their mechanical locking ability and self-retention system [1,2]. Our findings reinforce the notion that mechanical design, particularly conical locking systems like Morse taper, plays a pivotal role in reducing microleakage and microbial ingress under static in vitro conditions.

 

The amount of leakage that occurs between an implant and its abutment plays a fundamental role in causing peri-implant tissue inflammation. This research demonstrated that the Morse taper connection produced the least microleakage but the external hexagon connection indicated the highest microleakage results. Studies have established that Morse taper interfaces create a stronger seal preventing microorganisms from penetrating into the interface zone [3,4]. The susceptibility to develop microleakage is higher in external hexagon connections since they experience microgaps which cannot withstand occlusal loading forces [5,6].

 

The success rate of dental implants depends heavily on how well bacteria can penetrate their structure. The research findings demonstrated that bacterial infiltration levels were markedly lower in Morse taper connections relative to both internal hexagon and external hexagon connections. Past research has confirmed that Morse taper connections exhibit lower bacterial counts because their smaller microgap size makes it harder for microorganisms to invade [7,8]. The internal and external hexagon connections showed increased bacterial penetration rates that might lead to peri-implantitis and implant failure according to research findings [9,10].

 

Bacterial load tests confirm these results by measuring lower bacterial levels in Morse taper connections than in the other tested methods. Available literature supports the conclusion that bacterial leakage decreases when using conical connections compared to hexagonal connections [11,12]. The Morse taper connection blocks bacterial contamination through its friction-fit concept that reduces bacterial accumulation sites [13].

 

Several scientific studies demonstrate how mechanical factors determine the extent of bacterial leakage that develops between implant-abutment joints. Researchers have shown that hexagonal connection interfaces produce movement that allows bacteria to penetrate when stress is applied in functionally loaded conditions [14]. Morse taper connections present better mechanical stability over other systems by lowering the threat of bacterial microleakage and implant complications [15]. The method of abutment tightening plays a crucial role in developing microleakage between components. Calibrated wrenches delivered torque according to manufacturer guidelines yet actual microgap alterations might occur from minor torque-related changes. The leakage in hexagonal systems tends to get worse due to improper prosthetic fit and mechanical instability that occurs when patients perform chewing functions. The explored variables must be included in upcoming dynamic testing protocols because they mirror essential clinic aspects although not employed during this study.

 

The testing environment used outside the body fails to duplicate all aspects of the oral ecosystem since it does not include pH variations or masticatory stress or temperature changes. The findings from this study could be limited in their practical clinical application due to the absence of fatigue loading as well as real-time prosthetic use. Studies must include the essential variable components in future research designs along with complete analysis of long-term structural change in vivo.

 

Future Recommendations

Clinical studies are necessary to evaluate how Morse taper implants perform under cyclic loading and in patients with compromised immunity, diabetes, or poor oral hygiene. Moreover, investigations into the effect of torque variability, abutment fit, and advanced antibacterial coatings on microleakage should be pursued to enhance our understanding of connection design optimization.

CONCLUSIONS

The research data shows that implant-abutment connections designed with Morse taper exhibit better performance against microleakage and bacterial penetration. The clinical implementation of Morse taper designs now has evidence showing their effectiveness to prevent peri-implantitis and promote extended implant service. The use of Morse taper systems stands as an excellent choice for clinical applications which combine elevated risk of infection with extended prosthesis life requirements. Clinical experiments should be performed with dynamic loading in combination with varied patient groups including smokers and diabetics to validate the in vitro research outcomes.

REFERENCES

1. Donlan, Rodney M., and J. William Costerton. “Biofilms: Survival mechanisms of clinically relevant microorganisms.” Clinical Microbiology Reviews, vol. 15, no. 2, April 2002, pp. 167-193. http://dx.doi.org/10.1128/cmr.15.2.167-193.2002.

2. Filho, Antônio Pedro Ricomini, et al. “Preload loss and bacterial penetration on different implant-abutment connection systems.” Brazilian Dental Journal, vol. 21, no. 2, 2010, pp. 123-129. http://dx.doi.org/10.1590/s0103-64402010000200006.

3. Relhan, Nidhi, et al. “Endophthalmitis: Then and now.” American Journal of Ophthalmology, vol. 187, March 2018, pp. 20-27. http://dx.doi.org/10.1016/j.ajo.2017.11.021.

4. Baj, A., et al. “Bacterial colonization of the implant-abutment interface of conical connection with an internal octagon: an in vitro study using real-time PCR.” Journal of Biological Regulators and Homeostatic Agents, vol. 31, no. 2, 2017, pp. 163-168. https://pubmed.ncbi.nlm. nih.gov/28691468/.

5. Baj, A., et al. “Cone-morse implant connection system significantly reduces bacterial leakage between implant and abutment: an in vitro study.” Journal of Biological Regulators and Homeostatic Agents, vol. 31, no. 2, 2017, pp. 203-208. https://pubmed.ncbi.nlm.nih.gov/28691474/

6. Zhou, Cailing, et al. “Amphiphilic nano-swords for direct penetration and eradication of pathogenic bacterial biofilms.” ACS Applied Materials & Interfaces, vol. 15, no. 16, April 2023, pp. 20458-20473. http://dx.doi.org/10.1021/acsami.3c03091.

7. Mao, Zhen, et al. “Microleakage along the implant–abutment interface: A systematic review and meta-analysis of in vitro studies.” International Journal of Implant Dentistry, vol. 9, no. 1, September 2023. http://dx.doi.org/10.1186/s40729-023-00494-y.

8. Ranieri, Rogerio, et al. “The bacterial sealing capacity of morse taper implant–abutment systems in vitro.” Journal of Periodontology, vol. 86, no. 5, May 2015, pp. 696-702. http://dx.doi.org/10.1902/jop.2015.140623.

9. Mencio, F., et al. “Salivary bacterial leakage into implant-abutment connections: preliminary results of an in vitro study.” European Review for Medical and Pharmacological Sciences, vol. 20, no.12, June 2016, pp. 2476-2483. https://pubmed.ncbi.nlm.nih.gov/27383295/.

10. Nascimento, C. do, et al. “Bacterial leakage along the implant–abutment interface of premachined or cast components.” International Journal of Oral and Maxillofacial Surgery, vol. 37, no. 2, February 2008, pp. 177-180. http://dx.doi.org/10.1016/j.ijom.2007.07.026.

11. Liu, Xiaohui, et al. “Breaching bacterial biofilm barriers: Efficient combinatorial theranostics for multidrug-resistant bacterial biofilms with a novel penetration-enhanced aiegen probe.” ACS Applied Materials & Interfaces, vol. 14, no. 37, September 2022, pp. 41671-41683. http://dx.doi.org/10.1021/acsami.2c07378.

12. Assenza, Bartolomeo, et al. “Bacterial leakage in implants with different implant–abutment connections: An in vitro study.” Journal of Periodontology, vol. 83, no. 4, April 2012, pp. 491-497. http://dx.doi.org/10.1902/jop.2011.110320.

13. Yu, Z., et al. “[Establishment of animal model of bacterial microleakage at implant-abutment interface].” Zhonghua Kou Qiang Yi Xue Za Zhi, vol. 55, no.5, May 2020, pp. 337-342. https://pubmed.ncbi.nlm.nih.gov/36083296/.

14. Chen, Zhaowei, et al. “Enzyme mimicry for combating bacteria and biofilms.” Accounts of Chemical Research, vol. 51, no. 3, February 2018, pp. 789-799. http://dx.doi.org/10.1021/acs.accounts.8b00011.

15. Khorshidi, Hooman, et al. “In vitroEvaluation of bacterial leakage at implant-abutment connection: An 11-degree morse taper compared to a butt joint connection.” International Journal of Biomaterials, vol. 2016, 2016, pp. 1-5. http://dx.doi.org/10.1155/2016/8527849.

Recommended Articles
Research Article

Evaluation of Histological Changes Resulting from the Effect of the Drug Pregnyl on the Histological Structure of The Lungs in the White Mouse Mus Musculus

...
Published: 05/08/2025
pdf Download PDF
Research Article

Molecular Diagnosis of Mycoplasma Pneumonia Isolated from Hospitals in Hafer Al Batin, Saudi Arabia

Published: 05/08/2025
pdf Download PDF
Research Article

Drug Performance Indicators, Cons and Pros Upon Extended use or Misuse of Proton Pump Inhibitors

...
Published: 05/08/2025
pdf Download PDF
Research Article

Microbiological Assessment of Food Safety and Hygiene Practices Among Street Food Vendors at Khartoum Central Market

Published: 05/08/2025
pdf Download PDF
Copyright © Journal of Pioneering Medical Sciences until unless otherwise.